Service manual
2. TROUBLESHOOTING
2.26 Service Manual CNS-9701
Hard Disk Replacement Flow
When you order a new hard disk for replacement, the new hard disk has the OS
(Windows 2000) and central monitor system program installed at the factory.
Therefore, you do not have to install these. Replace the old hard disk with the new
one according to the following procedure.
1. Write down the following settings.
- IP address of the personal computer used as the central monitor main unit.
- The central monitor settings which are different from the factory default
settings.
- IP address of the network printer
2. Save the printer driver software on a storage media such as floppy disk or CD-
R if you use a printer which is not recommended by us.
3. Replace the old hard disk with the new one according to the “Replacing the
Hard Disk” in Section 4 “Disassembly and Assembly”.
4. Connect the display unit, keyboard, mouse and protection key to the central
monitor main unit. Refer to “Connection Diagram of Central Monitor” in
Section 1.
5. Turn on the peripheral units and main unit. The main unit automatically
recognizes the hardware components. The hardware information stored in the
new hard disk changes to the actual hardware information. The central
monitor system program automatically runs.
6. Shut down the central monitor according to the shutdown procedure in the
operator’s manual.
7. If a restart request dialog box appears on the screen, restart Windows 2000.
8. Check that there is no restart request dialog box on the screen. If the restart
request dialog box is still there, repeat the restart of Windows 2000 until the
restart request dialog box does not appear. The number of restart request
dialog box appearances depends on the hardware composition of the central
monitor.
9. Run the central monitor system program.
10. Set the IP address on the central monitor. Refer to “Setting the IP Address on
Each Instrument” in Section 8 “Network Settings”.
11. Connect the central monitor to the network. Refer to “Network and System”
installation guide. On the central monitor, check that each instrument is
connected to the network. Refer to “Checking the Network Connection on the
Central Monitor” in Section 8 “Network Settings”.