2. TROUBLESHOOTING
2.12 Service Manual CNS-9701
Overview Bed Screen
Parameter Setup
Trouble Possible Cause/Criteria Action
Some waveforms are not
displayed.
When the individual bed screen of
a patient is displayed on a central
monitor that is connected to a
multiple patient receiver or a signal
exchanger, some of the patient
waveforms cannot be displayed on
the Overview Bed of another
central monitor.
Display the All Beds screen at the central
monitor that saves the data of the bedside
monitor you want to display.
The desired bed cannot be
selected for the Overview Bed
screen.
The bed is an 8000 series bed.
To display data of an 8000 series bed on
this central monitor, a central monitor on
the network must be monitoring the bed.
The alarm for the overview bed
does not occur.
No bedside monitor is selected at
the Overview Bed screen.
Select a desired bed as an overview bed.
Cannot silence an overview bed
alarm.
An overview bed alarm of an
unintended bed occurs.
An overview bed alarm cannot be
silenced at the central monitor.
To stop overview bed alarm, stop
monitoring any overview bed by touching
“Stop monitoring” on the Overview Bed
Menu screen.
Trouble Possible Cause/Criteria Action
Available settings differ among
bedside monitors.
Available settings depend on the
bedside monitor model and
connection method.
---