Service manual
2. TROUBLESHOOTING
Service Manual CNS-9701 2.11
Report Window
Recording
Trouble Possible Cause/Criteria Action
Report printing starts
automatically.
The report printing mode is set to
“Auto”.
Set the mode to “Manual”.
Trouble Possible Cause/Criteria Action
In the following recordings,
waveforms that you do not want
to record are recorded.
- Manual recording
- Periodic recording
- Alarm recording
- Call recording
Waveforms displayed on the All
Beds screen are recorded.
---
In the following recording,
waveforms that you do not want
to record are recorded.
- Dual waveform recording
Recorded waveform is not selected
on the Recording window of the
Setup.
Select the desired waveform on the
Recording window of the Setup.
In the following recording,
waveforms that you do not want
to record are recorded.
- Remote delayed waveform
recording
Recorded waveform is selected at
the bedside monitor.
---
Alarm recording is not
performed for a particular
arrhythmia.
Recall file creation for the
arrhythmia is set to OFF.
Turn the setting to ON.
Vital sign alarm recording is not
performed.
Vital sign alarm recording is set to
OFF.
Set vital sign alarm recording to ON.
Arrhythmia alarm recording is set
to OFF.
Set arrhythmia alarm recording to ON.Arrhythmia alarm recording is
not performed.
Arrhythmia analysis is set to OFF
at the bedside monitor.
Turn arrhythmia analysis to ON at the
bedside monitor.
ST alarm recording is not
performed.
ST alarm recording is set to OFF. Set ST alarm recording to ON.
Periodic recording is not
performed.
Periodic recording is set to OFF on
the Recording window of the
Setup.
Set periodic recording to ON.
Call recording is not performed. Call recording is set to OFF on the
Recording window of the Setup.
Set call recording to ON.
Trend data are not printed for
the report printing.
Report settings are not correct. Set the correct settings on the Report
window.
Dual waveform recording is not
performed. (Only one waveform
is recorded/printed.)
On the Recording window of the
Setup, “Second Wave” at
“Recording Pattern” is set to
NONE.
Set a desired waveform for “Second
Wave” at “Recording Pattern”.
Waveform is not printed on the
screen copy.
Waveform cannot be printed on the
screen copy.
---