Service manual

2. TROUBLESHOOTING
2.10 Service Manual CNS-9701
Full Disclosure Window
ECG 12 Lead Analysis Window
Trouble Possible Cause/Criteria Action
The central monitor where the
bed’s data is saved is turned off.
Turn on the central monitor where the
bed’s data is saved.
The QP-973P Full Disclosure
Filing Program Kit is not installed.
Install the QP-973P Full Disclosure Filing
Program Kit.
Full disclosure waveform is not
displayed.
The QP-973P Full Disclosure
Filing Program Kit is not installed
in the central monitor where the
bed’s data is saved.
Choose a central monitor with the QP-
973P Full Disclosure Filing Program Kit
installed for the central monitor where the
bed’s data is saved.
The desired full disclosure
waveform is not displayed.
The full disclosure waveform is
not saved.
Choose the waveform for the stored
waveform on the Full Disclosure window.
The parameter is not measured. Measure the parameter.
The desired waveform cannot be
selected for a stored waveform.
This central monitor is not the
central monitor where the bed’s
data is saved.
This setting is only available at the central
monitor where the bed’s data is saved or
the BSS-9800 bedside station (central
mode).
The central monitor where the
bedsdataissavedwasturnedoff
during measurement.
Full disclosure waveform is not saved
when the central monitor where the bed’s
data is saved is turned off.
The waveforms that are selected
for the stored waveform are not
saved.
The currently measured lead/label
is different from the lead/label to
be saved.
Make the currently measured lead/label
match the lead/label to be saved.
Arrhythmia analysis is set to OFF
at the bedside monitor and/or the
central monitor.
Turn arrhythmia analysis to ON at the
bedside monitor and central monitor. You
cannot set color to an arrhythmia that does
not have color.
The arrhythmia alarm is set to OFF
at the bedside monitor and/or the
central monitor.
Turn the arrhythmia alarm to ON at the
bedside monitor and central monitor. You
cannot set color to an arrhythmia that does
not have color.
On the Full Disclosure window,
the arrhythmia waveform is not
indicated with the preset color.
Noise interference
Poor electrode attachment
condition
You cannot set color to an arrhythmia that
does not have color.
Trouble Possible Cause/Criteria Action
The QP-974P 12 Lead ECG
Interpretation Filing Program Kit is
not installed.
Install the QP-974P 12 Lead ECG
Interpretation Filing Program Kit.
ECG 12 Lead analysis files are
not displayed.
The bedside monitor cannot output
ECG 12 lead analysis files to a
network.
The central monitor can display ECG 12
lead analysis files of the BSM-4100 series
bedside monitor only when the bedside
monitor version is 03-02 or later.