Service manual

2. TROUBLESHOOTING
Service Manual CNS-9701 2.9
Arrhythmia Recall Window
ST Recall Window
Trouble Possible Cause/Criteria Action
Arrhythmia recall files are not
created.
The arrhythmia analysis is set to
OFF.
Turn the arrhythmia analysis to ON.
Arrhythmia recall files for a
particular arrhythmia are not
created
Recall file creation for the
arrhythmia is set to OFF.
Turn the setting to ON.
Arrhythmia recall files for
unnecessary arrhythmias are
created.
Recall file creation for the
arrhythmia is set to ON.
Turn the setting to OFF.
The number of arrhythmia recall
files for one bed exceeded the limit
and older files are deleted.
Arrhythmia recall files are
deleted.
You touch “Delete” when several
files have a deletion mark.
Deleted files cannot be recovered.
The bedside monitor is turned off. Turn the bedside monitor or multiple
patient receiver on.
The central monitor where the
bed’s data is saved is turned off.
Turn on the central monitor where the
bed’s data is saved.
Arrhythmia recall files are not
displayed.
The arrhythmia types to be
displayed on the screen are not
selected.
Select the arrhythmias to be displayed on
the screen.
Arrhythmia recall files are not
saved.
The central monitor where the
bed’s data is saved was turned off
during measurement.
Arrhythmia recall files are not created
when the central monitor where the bed’s
data is saved is turned off.
When the multi waveform
display is selected, the desired
waveforms cannot be displayed.
Stored waveforms and displayed
waveforms are not selected for full
disclosure waveform.
Waveforms displayed for the multi
waveform display are synchronized with
the displayed waveform settings on the
Full Disclosure window. Set at “Stored
Waves” and “Displayed Waves” on the
Full Disclosure window.
Multi waveform cannot be
displayed.
The QP-973P Full Disclosure
Filing Program Kit is not installed.
Install the QP-973P Full Disclosure Filing
Program Kit.
Trouble Possible Cause/Criteria Action
ST recall files for 72 hours
cannot be saved.
The interval of ST recall file
creation is set to 30 seconds.
With the creation interval of 30 seconds,
the total number of files exceeds the limit.
To save recall file for 72 hours, set the
creation interval to 1 min or more.
The central monitor where the
bed’s data is saved is turned off.
Turn on the central monitor where the
bed’s data is saved.
ST recall files are not saved.
The bedside monitor cannot output
ST recall files to a network.
The central monitor cannot save or display
ST recall files of the following bedside
monitors:
BSM-4100 series bedside monitor
whose version is not 02-09 or later
8000 series bed that is a monitored bed
of a CNS-9301/9302 central monitor.