Service manual

2. TROUBLESHOOTING
2.8 Service Manual CNS-9701
Trend Window
Tabular Trend Window
Hemodynamics List Window
Trouble Possible Cause/Criteria Action
Trend data is not displayed. The central monitor where the
bed’s data is saved is turned off.
Turn on the central monitor where the
bed’s data is saved.
Trend data is not saved. The central monitor where the
bedsdataissavedwasturnedoff
during measurement.
Trend data is not created when the central
monitor where the bed’s data is saved is
turned off.
Trouble Possible Cause/Criteria Action
Tabular trend data is not
displayed.
The central monitor where the
bed’s data is saved is turned off.
Turn on the central monitor where the
bed’s data is saved.
Cannot select a parameter you
want to add to the tabular trend
The parameter is not registered at
the Parameters window of the
System Setup.
Register the parameter at the Parameters
window of the System Setup.
NIBPnumericdataisnot
displayed.
When an NIBP measurement is
delayed or remeasured and its
measurement timing does not
match the selected tabular trend
interval, that NIBP data will not
appear on the tabular trend.
Set the tabular trend display interval to 1
min. The NIBP tabular trend and all NIBP
measurements are displayed.
Tabular trend data is not saved. The central monitor where the
bedsdataissavedwasturnedoff
during measurement.
Tabular trend data is not created when the
central monitor where the bed’s data is
saved is turned off.
Trouble Possible Cause/Criteria Action
Hemodynamics data is not
displayed.
The central monitor where the
bed’s data is saved is turned off.
Turn on the central monitor where the
bed’s data is saved.
The central monitor where the
bedsdataissavedwasturnedoff
during measurement.
Hemodynamics data is not created when
the central monitor where the bed’s data is
saved is turned off.
Hemodynamics data is not
saved.
The bedside monitor cannot output
hemodynamics data to the
network.
Hemodynamics data that can be reviewed
at the central monitor is data created at the
BSM-4100 series bedside monitor version
02-09 or later.