Service manual

2. TROUBLESHOOTING
2.6 Service Manual CNS-9701
Admitting/Discharging
Trouble Possible Cause/Criteria Action
Cannot transfer a patient’s data.
Cannot change the receiving
channel.
You are trying to do this operation
on a central monitor where the
patient’s data is not saved.
Do this on the central monitor where the
patient’s data is saved.
You are trying to do this operation
on a central monitor where the
patient’s data is not saved.
Do this on the central monitor where the
patient’s data is saved.
Cannot admit/discharge a
patient.
The bedside monitor is turned off. Turn on the bedside monitor.
Cannot select a bedside monitor
as a transfer destination bedside
monitor.
The bedside monitor is turned off. Turn on the bedside monitor.
The patient’s data before patient
transfer is lost.
The patient was transferred
between bedside monitors in
different CNS groups more than
once.
To save the patient’s data in the first
central monitor, do not transfer the patient
to a bedside monitor in a third CNS group
after the patient is transferred to a bedside
monitor in a second CNS group.
The patient’s data before patient
transfer cannot be reviewed.
The central monitor where the
patient’s data is saved is turned off.
Turn on the central monitor where the
patient’s data before the patient transfer is
saved.
The message “Transferring” is
displayed on the screen but it
does not change to
“Completed”.
All patient’s data is lost after
patient transfer.
During patient transfer, the central
monitor was turned off. Or, there
was trouble on the network and the
transfer failed.
Admit the patient at the destination
bedside monitor. The lost data cannot be
restored.
ST recall data is lost after
patient transfer.
You transferred an 8000 series bed
that was a monitored bed of a
CNS-9301/9302 central monitor.
You cannot transfer ST recall data of an
8000 series bed that was a monitored bed
of a CNS-9301/9302 central monitor.