Service manual

Service Manual CNS-9701 1.13
1. GENERAL
Arrhythmia: ASYSTOLE, V FIB, V TACHY, EXT TACHY,
EXT BRADY, VPC RUN, COUPLET, MULTI
FORM, EARLY VPC, BIGEMINY, TACHY,
BRADY, PROLONGED RR, FREQ. VPC
Apnea
Alarm display Highlighted numeric display, highlighted message for
arrhythmia
Alarm occurrence Alarm occurs when any one of the bedside monitors that the
central monitor are monitoring generates alarm.
Alarm suspend Available (for beds connected via the QI-910R interface)
Alarm silence Available (except for beds connected via the QI-910R
interface)
Alarm recording Available
Trend parameters Depends on the connected bedside monitor or
transmitter
Trend display times 1, 8, 24, 72 h
Trend display formats Trendgraph and tabular trend
Events related to trendgraph ASYSTOLE, V FIB, V TACHY, VPC RUN,
COUPLET, Off/Noise
ECG display sensitivity ×1/4, ×1/2, ×3/4, ×1, ×1.5, ×2, ×4
Respiration curve sensitivity ×1/4, ×1/2, ×3/4, ×1, ×1.5, ×2, ×4
BP display scale (mmHg) 0-20, 0-50, 0-100, 0-200, 50-200, 50-250, 0-300
mmHg
Displays user-selectable vital signs, up to 12 ECG waveforms, trend data, alarm
events, and status messages associated with the selected overview bed.
The overview bed can be any bed in the network that the central monitor is not
continuously monitoring.
Alarm setting of bedside monitors and multiple patient receivers
Individual setting of bedside monitors and multiple patient receivers
Channel setting of multiple patient receivers
Saves 72 hours of full disclosure waveform data for up to six waveforms and
displays them on the Full Disclosure window.
(When the QP-973P Full Disclosure Filing Program Kit is installed.)
Trend
Arrythmia recall file
Hemodynamic data list
ST recall
Trend
Waveform Sensitivity
Overview
Remote Setting
Full Disclosure
File Saving