Service manual

1.12 Service Manual CNS-9701
1. GENERAL
Number of patients on the display 16 patients max
With overview bed function, 16 patients plus
one other patient.
Number of transmitters 16 transmitters max
Display type 18 inch color TFT
Waveform display method Non-fade, fixed method
Number of waveform traces Single display: 24 traces max (On the All
Beds screen, displaying 12
patients)
Dual display: 32 traces max (On the All
Beds screen, displaying 16
patients)
Sweep speed 25 mm/s, 50 mm/s, 6.25 mm/s (respiration
measurement)
Sweep time All Beds screen (8 patients): 4.48 s
All Beds screen (4 patients): 11.56 s
Individual Bed screen: 10.28 s
ECG1 - 12, IBP (1- 8), respiration wave, EEG, CO
2
, external input, pulse (SpO
2
),
Flow/Paw, anesthetic gas (O
2
, CO
2
, N
2
O, Agent), others depending on the
connected bedside monitor or transmitter
Heart rate, VPC rate, respiration rate, pulse rate, IBP (systolic, diastolic, mean),
NIBP (systolic, diastolic, mean), temperature or T or blood temperature, EtCO
2
,
tcPO
2
, tcPCO
2
, SpO
2
, ST level, FiO
2
, CO, CCO, respirator, anesthetic gas, BIS,
Sv
_
O
2
, Flow/Paw, N
2
O, O
2
,
agent, others depending on the connected bedside
monitor or transmitter
Synchronization source ECG
Alarm sound Crisis, Warning, Advisory
Sound volume changeable for Heart rate sync mark and sound
Alarm decision is done at each bedside monitor and multiple patient receiver. The
central monitor only displays the alarm and generates sound by receiving the alarm
information from bedside monitors and multiple patient receivers.
Alarm type Crisis, Warning, Advisory
Alarm items Vital Sign: Heart rate, VPC rate, Respiration rate, Pulse
rate, ST level, BP (sys, dia, mean), NIBP (sys,
dia, mean), temperature or T or blood
temperature, EtCO
2
, tcPO
2
, tcPCO
2
, SpO
2
, FiO
2
,
CCO, others depending on the connected
bedside monitor or transmitter
Specification
Display
Waveform Display Items
Alphanumeric Display
Items
Sound
Alarm Function