Service manual
Service Manual CNS-9701 1.9
1. GENERAL
Temporary Discharge
WARNING
When a temporarily discharged patient comes back, resume
monitoring for the patient. During temporary discharge, there is no
display of measurement values, alarm function and data saving.
Patient Transfer
CAUTION
••
••
• Keep the original bedside monitor and the destination bedside
monitor power on and connected to the network until the transfer
completes. Otherwise transfer fails and the data is lost. After
transfer, confirm that the data was correctly transferred to the
destination bedside monitor.
••
••
• When the patient is transferred more than once between different
central monitors in the network, the data saved on the original
central monitor is deleted.
Alarm
WARNING
••
••
• When the Alarm Suspend key on the transmitter is pressed, all alarm
functions for the patient are suspended.
••
••
• Even if the alarm is set to ON, there is no arrhythmia alarm when
arrhythmia analysis is turned OFF at the bedside monitor and/or
central monitor.
••
••
• Before you start monitoring a new patient, confirm alarm settings.
Alarm settings return to Alarm Master 1 settings when a bed is
discharged or a receiving channel is changed.
••
••
• Securely connect the sound cable. There is no sound if the cable
comes off.
CAUTION
••
••
• When the upper or lower alarm limit is set to OFF, there will be no
upper or lower alarm for that limit. When it is set to OFF, frequently
observe the patient.
••
••
• Alarm does not occur for an arrhythmia whose alarm is set to OFF.
There is no indication that the alarm is set to OFF. Be careful when
you set the alarm to OFF.