Service manual
Service Manual CNS-9701 1.7
1. GENERAL
Network
WARNING
••
••
• Connect the central monitor to the network as specified in the
Network and System Installation Guide. Otherwise the patient and
operator may get electrical shock or other injury. For connecting the
network, contact your Nihon Kohden distributor or representatives.
••
••
• Check the software version number of the monitor before connecting
it to the network. Different software versions have different
communication methods. When there is more than one
communication method in the network, communication may
malfunction.
••
••
• Install the external instruments including the printer and hubs
outside the patient environment (IEC 60601-1-1 2.204). If they are
installed inside the patient environment, the patient or operator may
receive electrical shock. For installation, contact your Nihon Kohden
distributor or representatives.
CAUTION
••
••
• Network settings and Windows 2000 system settings must only be
set by a qualified service personnel. Contact your Nihon Kohden
distributor or representatives.
••
••
• The network must be managed by the network administrator. Make
sure that each monitor in the network has a different IP address.
Otherwise, data communication cannot be performed properly and it
causes incorrect monitoring. When adding a monitor to an already
operating network, set the IP address on the monitor before
connecting the monitor to the network.