Service manual
8. NETWORK SETTINGS
Service Manual CNS-9701 8.17
One bedside monitor can be monitored on several central monitors, but the data
can only be stored on one central monitor. When you register a monitored bed on
the Monitor Settings of the System Setup, you will be asked if you want the central
monitor to store the bed’s data. Availability of setting and operations depends on
whether or not the bed’s data is saved in the central monitor or another central
monitor. Refer to the CNS-9701 operator’s manual.
A central monitor can monitor and store data of beds in a different group.
However, it is recommended to monitor and store data of the beds in the same
group as the central monitor so that data will not be mixed or lost.
Selecting the Monitored Beds
ICU-001
BSM
ICU-002
BSM
ICU-016
BSM
ICU-003
BSM
CNS-ICU2
CNS
CNS-ICU1
CNS
ICU
CCU-001
BSM
CCU-002
BSM
CCU-008
BSM
CCU-003
BSM
CNS-CCU1
CNS
CCU
Group name
Group name
Network
CNS name
Bed name
The CCU central monitor can monitor and store data of the beds in
the ICU group. However, it is recommended to monitor and store
data of the CCU beds in the CCU group so that data will not be lost.
Bed name
CNS name
CAUTION
The data of a bedside monitor cannot be stored on more than one
central monitor. If the data of the same bedside monitor is stored on
two central monitors, data communication cannot be properly
performed or the data may be lost. Make sure that the bedside monitor
data is saved only on one central monitor.