8. NETWORK SETTINGS
Service Manual CNS-9701 8.15
Assign an instrument name to the central monitor, beds on the multiple patient
receiver and bedside monitors on the network.
Central monitor:
Exit the central monitor system and assign a name in Windows 2000.
Beds on the ORG-9200 multiple patient receiver:
First register the bed as a monitored bed on the Monitor Setting window of the
System Setup. Then set the name for the bed. For the procedure, refer to the CNS-
9701 operator’s manual.
Bedside monitors:
Assign the bed name to bedside monitors on the SYSTEM SETUP screen on each
bedside monitor. Refer to the bedside monitor operator’s manual.
Assigning a CNS Name to a Central Monitor
1. Shut down the central monitor system.
2. Select “Start” - ”Setting” - ”Control Panel”. The “Control Panel” dialog box
opens.
3. Click the “System” icon. The “System Properties” dialog box opens.
4. Click “Network Identification”.
5. Click the “Properties” button. The “Identification changes” dialog box opens.
6. Enter the name in the “Computer name” column.
7. Click the “OK” button.
8. Restart Windows 2000. Restart is necessary to enable the new settings.
Up to 16 group names can be registered for one network. If you do not divide
instruments on a network into groups, registering group names is unnecessary.
Registration can be done on the Name Registration window of the System Setup.
Refer to Section 4 of the CNS-9701 operator’s manual. This setting affects the
entire network because group names are shared in the entire network.
Assigning a Bed Name/
CNS Name
Registering Group Names
in the Central Monitor