Service manual
8. NETWORK SETTINGS
8.14 Service Manual CNS-9701
To identify the monitors, each monitor has an instrument name (bed name/CNS
name) and a group name. Every bed and every central monitor belongs to one of
up to 16 groups. A local bed is assigned to the same group as the central monitor.
Groups are provided to simplify bed selection in large hospitals. It is
recommended to assign a different bed name to each bed and central monitor so
that it is easy to identify a bed.
Assigning the Group Name and Bed Name
ICU-001
BSM
ICU-002
BSM
ICU-016
BSM
ICU-003
BSM
CNS-ICU2
CNS
CNS-ICU1
CNS
ICU
CCU-001
BSM
CCU-002
BSM
CCU-008
BSM
CCU-003
BSM
CNS-CCU1
CNS
CCU
Group name
Group name
Network
CNS name
Bed name
The CCU central monitor can monitor and store data of the beds in
the ICU group. However, it is recommended to monitor and store
data of the CCU beds in the CCU group so that data will not be lost.
Bed name
CNS name
CAUTION
When the monitor is connected to a central network, set the bed name/
CNS name and group name on the monitor to correctly identify the bed
on the central monitor. Otherwise the monitor cannot be managed
properly.