Service manual
Instrument Functions WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
65Operating Manual 1176.7649.02 ─ 04
For IEEE 802.11b and g (single carrier), the PVT results are displayed as percentage
values of the reference power. The reference can be set to either the max or mean
power of the PPDU. For both rising and falling edges two time lines are displayed,
which mark the points 10 % and 90 % of the reference power. The time between these
two points is compared against the limits specified for the rising and falling edges.
For further details see also chapter 3.8.3, "Rise/Fall Time Measurement", on page 50
Remote command:
CONFigure:BURSt:PVT[:IMMediate] on page 148
Full Burst ← PVT
Displays the PVT results in a single graph with all PPDU data being displayed.
For further details refer to the "PVT" on page 64 softkey.
Remote command:
CONFigure:BURSt:PVT:SELect on page 149
Rising & Falling ← PVT
Displays the PVT results in two separate graphs, the left hand side showing the rising
edge and the right hand side showing the falling edge.
Softkeys of the WLAN TX Menu