Service manual
WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
54Operating Manual 1176.7649.02 ─ 04
Setting Description Restrictions
Standard Selected measurement standard
PPDU/MCS
Index/GI
The PPDU Type, MCS Index and Guard Interval
used for the analysis of the signal is displayed.
Depending on the Demod Settings, these values are
either automatically detected from the signal or the
user settings are applied.
Burst x of y (z) In case statistic over bursts is switched on (Overall
Burst Count), x bursts of totally required y (No of
Bursts to Analyze) bursts have been analyzed so
far. The value z gives the number of analyzed
bursts by the last update of the statistic.
3.9.2 Result Summary List
The result summary list shows the overall measurement results and provides limit
checking for result values in accordance with the selected standard. Result values
which are within the limit as specified by the standard are displayed in green. Result
values which are outside of the limits specified by the standard are displayed in red
with a '*' to the left. Results which have no limits specified by the standard are dis-
played in white. Limit values are displayed in white (not bold) and can be modified, if
focused, via the keypad. To reset the limit values to the values specified in the stand-
ard, use the "Lines" menu ( chapter 4.8, "Softkeys of the Lines Menu – LINES key",
on page 113).
The results displayed in this list are for the entire measurement. If a specific number of
bursts have been requested which requires more than one sweep, the result summary
list is updated at the end of each sweep. The number of bursts measured and the num-
ber of bursts requested are displayed to show the progress through the measurement.
The Min/Mean/Max columns show the minimum, mean or maximum values of the burst
results.
For details on the displayed measurement results see chapter 3.8, "Measurement
Result Types", on page 43.
Measurement Settings and Result Displays