Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
220Operating Manual 1176.7649.02 ─ 04
TRIGger[:SEQuence]:LEVel:POWer:AUTO <State>
This command specifies whether or not an automatic power trigger level calculation is
performed before each main measurement. The setting of this command is ignored if
the setting for the TRIGger[:SEQuence]:MODE on page 218 command is not
POWer.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
TRIG:LEV:POW:AUTO ON
Specifies that an automatic power trigger level calculation
should be performed before the start of each main measure-
ment.
Mode: WLAN
Manual operation: See "Auto Lvl " on page 91
5.18 UNIT Subsystem (K91)
UNIT:EVM.....................................................................................................................220
UNIT:GIMBalance..........................................................................................................220
UNIT:PREamble............................................................................................................ 221
UNIT:EVM <Unit>
This command specifies the units for EVM results.
Parameters:
<Unit> DB | PCT
DB
EVM results returned in dB
PCT
EVM results returned in %
*RST: DB
Example:
UNIT:EVM PCT
EVM results to be returned in %.
Mode: WLAN
UNIT:GIMBalance <Unit>
This command specifies the units for gain imbalance results.
UNIT Subsystem (K91)