Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
219Operating Manual 1176.7649.02 ─ 04
Parameters:
<Mode> IMMediate | EXTernal | POWer | PSENsor | RFPower
IMMediate
No triggering is performed. This corresponds to the Free Run
trigger mode.
EXTernal
The next measurement is triggered by the signal at the external
trigger input e.g. a gated trigger.
POWer
The next measurement is triggered by signals outside the mea-
surement channel.
PSEN
The next measurement is triggered by the external power sensor
(requires R&S FSV-K9 option).
RFPower
The next measurement is triggered by the first intermediate fre-
quency of the RF signal.
TUNit
If activated, the next measurement is triggered by the signal at
the external trigger input connected to the R&S
®
FS-Z11 trigger
unit. This allows for all connected analyzers to be synchronized
for MIMO measurements.
For details see chapter 3.6.1, "Trigger Synchronization Using an
R&S
®
FS-Z11 Trigger Unit", on page 37.
*RST:
IMMediate
Example:
TRIG:MODE IMM
No triggering is performed.
Mode: WLAN
Manual operation: See "Trigger Mode " on page 90
TRIGger[:SEQuence]:LEVel:POWer <Level>
This command sets the level of the input signal for which triggering will occur.
Parameters:
<Level> Range: -50 to 20
*RST: -20 DBM
Default unit: dBm
Example:
TRIG:MODE POW
Sets the external trigger mode.
TRIG:LEV:POW 10 DBM
Sets the level to 10 dBm for RF measurement.
Mode: WLAN
Manual operation: See "Power Level " on page 91
TRIGger Subsystem (WLAN)