Service manual
Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
217Operating Manual 1176.7649.02 ─ 04
is therefore a fixed value. Only an array of Y data is returned. TRACE1 is used for
these measurement results.
Supported data formats (see FORMat[:DATA] on page 176): ASCii|REAL
TRACE1 Clear write values
TRACE2 Max hold values
5.17 TRIGger Subsystem (WLAN)
The trigger subsystem is used to synchronize device action(s) with events.
TRIGger[:SEQuence]:HOLDoff........................................................................................217
TRIGger<n>[:SEQuence]:IFPower:HOLDoff..................................................................... 217
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis.................................................................218
TRIGger<n>[:SEQuence]:LEVel[:EXTernal]...................................................................... 218
TRIGger[:SEQuence]:MODE...........................................................................................218
TRIGger[:SEQuence]:LEVel:POWer................................................................................ 219
TRIGger[:SEQuence]:LEVel:POWer:AUTO.......................................................................220
TRIGger[:SEQuence]:HOLDoff <Delay>
This command defines the length of the trigger delay. A negative delay time (pretrig-
ger) can be set in zero span only.
Parameters:
<Delay> Range: -3.25 to 837.33
*RST: 0 s
Default unit: ms
Example:
TRIG:HOLD 500us
A holdoff period of 500 μs is used after the trigger condition has
been met.
Usage: SCPI confirmed
Mode: WLAN
Manual operation: See "Trigger Offset " on page 90
TRIGger<n>[:SEQuence]:IFPower:HOLDoff <Value>
This command sets the holding time before the next IF power trigger event.
Suffix:
<n>
.
irrelevant
Parameters:
<Value> *RST: 150 ns
TRIGger Subsystem (WLAN)