Service manual
Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
197Operating Manual 1176.7649.02 ─ 04
However, for signals in which the PPDU power levels differ significantly, this option
should be disabled as otherwise some PPDUs may not be detected.
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
A coarse burst search is performed based on the power levels of
the input signal.
OFF | 0
No pre-evaluation is performed, the entire signal is processed.
*RST: 1
Manual operation: See "Power Interval Search (IEEE 802.11ac only)" on page 108
[SENSe:]FREQuency:CENTer <Frequency>
This command defines the center frequency (frequency domain) or measuring fre-
quency (time domain).
Parameters:
<Frequency> Range: 0 to fmax
*RST: fmax/2
Default unit: Hz
f
max
is specified in the data sheet. min span is 10 Hz
Example:
FREQ:CENT 100 MHz
Manual operation: See "Frequency " on page 88
[SENSe:]POWer:ACHannel:MODE <Mode>
This command sets the ACP measurement mode for the IEEE 802.11j standard to
either absolute or relative.
Parameters:
<Mode> ABS | REL
ABS
Absolute measurement
REL
Relative measurement
*RST: REL
Example:
POW:ACH:MODE ABS
Sets the ACP measurement to absolute mode
Mode: WLAN