Service manual
Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
167Operating Manual 1176.7649.02 ─ 04
Usage: Query only
SCPI confirmed
Mode: WLAN
Manual operation: See "Ref Level " on page 92
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel[:RF] <Level>
This command can be used to retrieve or set the current internal instrument reference
level for RF input used when performing measurements.
Suffix:
<n>
.
1 | 2
irrelevant
<t> 1...3
irrelevant
Parameters:
<Level> <numeric value> in dB
*RST: -5 dB
Example:
DISP:TRAC:Y:RLEV?
Returns the current RF reference level in use.
DISP:TRAC:Y:RLEV: -20
Sets the instrument reference level to -20.
Mode: WLAN
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:IQ <Level>
This command can be used to retrieve or set the current internal instrument reference
level for baseband input used when performing measurements.
Suffix:
<n>
.
1 | 2
irrelevant
<t> 1...3
irrelevant
Parameters:
<Level> <numeric value> in V
*RST: 1 V
Example:
DISP:TRAC:Y:IQ?
Returns the current baseband reference level in use.
DISP:TRAC:Y:RLEV:IQ 1
Sets the instrument reference level to 1.
Mode: WLAN
DISPlay Subsystem (WLAN)