Service manual

Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
148Operating Manual 1176.7649.02 ─ 04
Mode: WLAN
Manual operation: See "EVM vs Symbol/Carrier" on page 69
CONFigure:BURSt:EVM:ESYMbol[:IMMediate]
This remote control command configures the measurement type to be EVM vs Symbol.
After this command has been executed, the specified measurement will only be started
when the user issues the INITiate<n>[:IMMediate] command.
Example:
CONF:BURS:EVM:ESYM
Configures the EVM vs Symbol measurement type.
Mode: WLAN
Manual operation: See "EVM vs Symbol/Carrier" on page 69
CONFigure:
BURSt:PREamble[:IMMediate]
This remote control command configures the measurement type to be Phase or Fre-
quency vs Preamble. After this command has been executed, the specified measure-
ment will only be started when the user issues the INITiate<n>[:IMMediate]
command.
Example:
CONF:BURS:PRE
Configures the preamble measurement type.
Mode: WLAN
Manual operation: See "Error Frequency/Phase" on page 71
CONFigure:BURSt:PREamble:SELect <Mode>
This remote control command configures the interpretation of the preamble measure-
ment results.
Parameters:
<Mode> PHASe | FREQuency
Example:
CONF:BURS:PRE:SEL PHAS
Configures the Phase vs Preamble measurement type.
Mode: WLAN
Manual operation: See "Error Frequency/Phase" on page 71
CONFigure:BURSt:PVT[:IMMediate]
This remote control command configures the measurement type to be Power vs Time.
After this command has been executed, the specified measurement will only be started
when the user issues the INITiate<n>[:IMMediate] command.
Example:
CONF:BURS:PVT
Configures the Power vs Time measurement type.
CONFigure Subsystem (WLAN)