Service manual

Remote Commands for WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
120Operating Manual 1176.7649.02 ─ 04
The numeric values associated to MAXimum/MINimum/DEFault can be queried by
adding the corresponding keywords to the command. They must be entered following
the quotation mark.
Example:
SENSe:FREQuency:CENTer? MAXimum
Returns the maximum possible numeric value of the center frequency as result.
<arbitrary block program data>
This keyword is provided for commands the parameters of which consist of a binary
data block.
5.2 ABORt Subsystem
ABORt
This command aborts a current measurement and resets the trigger system.
Example:
ABOR;INIT:IMM
Mode: all
5.3 CALCulate:BURSt Subsystem (WLAN)
The CALCulate:BURSt subsystem checks the I/Q measurement results.
CALCulate<n>:BURSt[:IMMediate].................................................................................. 120
CALCulate<n>:BURSt[:IMMediate]
This command forces the I/Q measurement results to be recalculated according to the
current settings.
Suffix:
<n>
.
1...4
irrelevant
Example:
CALC1:BURS
Starts the recalculation of the IQ measurement results.
Usage: Event
Mode: WLAN
Manual operation: See "Analyze" on page 97
ABORt Subsystem