Service manual

Instrument Functions WLAN TX Measurements
R&S
®
FSV-K91/91n/91ac/91p
115Operating Manual 1176.7649.02 ─ 04
DigIConf ← EXIQ
Starts the optional R&S DigIConf application. This softkey is only available if the
optional software is installed.
To return to the R&S FSV application, press any key on the front panel. The applica-
tion is displayed with the "EXIQ" menu, regardless of which key was pressed.
For details on the R&S DigIConf application, see the "R&S®EX-IQ-BOX Digital Inter-
face Module R&S®DigIConf Software Operating Manual".
Note: If you close the R&S DigIConf window using the "Close" icon, the window is
minimized, not closed.
If you select the "File > Exit" menu item in the R&S DigIConf window, the application is
closed. Note that in this case the settings are lost and the EX-IQ-BOX functionality is
no longer available until you restart the application using the "DigIConf" softkey in the
R&S FSV once again.
Remote command:
Remote commands for the R&S DigIConf software always begin with SOURce:EBOX.
Such commands are passed on from the R&S FSV to the R&S DigIConf automatically
which then configures the R&S EX-IQ-BOX via the USB connection.
All remote commands available for configuration via the R&S DigIConf software are
described in the "R&S®EX-IQ-BOX Digital Interface Module R&S®DigIConf Software
Operating Manual".
Example 1:
SOURce:EBOX:*RST
SOURce:EBOX:*IDN?
Result:
"Rohde&Schwarz,DigIConf,02.05.436 Build 47"
Example 2:
SOURce:EBOX:USER:CLOCk:REFerence:FREQuency 5MHZ
Defines the frequency value of the reference clock.
Softkeys of the Input/Output Menu for WLAN Measurements