CHAPTER 8: MAINTENANCE
Maintenance of the equipment should be completed by qualified personnel at least
once a year.AnX Robotica Corp. will make available on request circuit diagrams,
component part lists, descriptions, calibration instructions, or other information that
will assist service personnel to repair the device. Do not attempt to make any
modifications to the device. Any modification during the actual service life require
re-evaluation of safety and EMC.
The following routine inspection and cleaning is required:
Power Inspection
Regularly check power supply of the System for damage, cracking or wear. Do not
use the equipment when power voltage is out of specified range for the equipment
120VAC, 60Hz.
8.1 Replacement of Fuse
There are two fuses with fast-acting and high breaking capacity features. Each
specified as 125VAC 6A, and the dimension is Φ5mm in diameter, 20mm in length.
The two fuses are located near the power input terminal of the equipment (see Figure
9-1). Please use a fuse removal tool to remove the fuse caps and replace the related
fuse. AnX Robotica recommends that the fuses by replaced annually.
Figure 9- 1: Schematic Diagram of fuse replacement
Warning
Replacing of the fuse can only be done by qualified personnel.