Stomach Capsule User Manual Page | 110
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
Relative humidity: ≤80%
Air pressure: 86 kPa~106 kPa
The environment should be free of dust, with no mechanical vibration, no
significant noise nearby and be free of power interference.
Within 5m area coverage of the installed equipment, there should be no
magnetic resonance imaging equipment or strong electromagnetic
interference
7.2 Installation and Adjustment Instruction
Installation, movement or modification of the location of the System should be
completed by a qualified maintenance technician or by AnX Robotica Corp personnel.
Warning
Do not move the equipment without authorization. If necessary, contact AnX
Robotica Corp.
7.3 Training
7.3.1 Equipment Operation Training
The equipment operator should be trained by AnX Robotica Corp. personnel before
start using the system. Your facility may designate a trainer to train additional
personnel, as applicable.
7.3.2 Clinic Use Training
It is strongly recommended by AnX Robotica that the equipment be used only by
well-trained physician or health professional with clinical operational experience with
this System.
Warning
The equipment operator should be well trained before performing clinical
operations with patients.