User's Manual

Table Of Contents
NaviCam
®
Stomach Capsule User Manual Page | 106
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
1 : Scanning indicator, Yellow;
: Power Switch;
2 : Working indicator, Green;
: Charging port;
3 : Infrared light source to activate
capsule;
: Power switch indicator, white;
4 : Keys;
: Label plate.
6.3 Usage
6.3.1 Turn-on the Capsule
Before starting the Capsule, press the “Power Switch” to activate the Locator. Once
the Locator is on, the working indicator will flash. Place the Capsule front case
opposite the start infrared light source of the Capsule. Press and hold the “INIT
Button” to align the open position of the Capsule. The Capsule LED will start
blinking after the Capsule is opened normally. The Capsule is activated as shown in
Figure 7-2.
Figure 7-2 Schematic Diagram of turning-on the Capsule
6.3.2 Locate the Capsule
To detect the location of the capsule in a patient using the Locator, perform the
following steps:
1. Turn on: Hold the Locator upright, start the Locator by pressing “Power Switch”
in the open space without metal objects nearby. Once the Locator has started, the
working indicator will flash at a frequency of once per second, and the scanning
indicator will be off.
2. Locating: Use the Locator to move gently in the detection area. Once the Locator
is near the capsule, the scanning indicator will light up. When the scanning
indicator stays on, it indicates that the capsule is right below the scanning