User's Manual
Table Of Contents
- Chapter 1:INTRODUCTION
- Chapter 2:INDICATIONS, CONTRAINDICATIONS AND WARNINGS
- Chapter 3:WIRING AND CONTROL PANEL OF CONSOLE
- Chapter 4:OPERATION INSTRUCTION FOR ESNAVI SOFTWARE
- 4.1Installing Software
- 4.2Uninstalling Software
- 4.3Basic Operation of Software
- Chapter 5:USAGE OF CAPSULE
- Chapter 6:USAGE OF DATA RECORDER
- Chapter 7:INSTALLATION AND TRAINING
- Chapter 8:MAINTENANCE
- Chapter 9:WARNINGS, CAUTIONS AND TROUBLESHOOTING
- 9.1ESNavi Error Messages
- 9.1.1.1.1.Check if camera is connected firmly
- 9.1.1.1.2.Equipment initialization error
- 9.1.1.1.3.Port "COM7" does not exist
- 9.1.1.1.4.Video monitoring connection fails
- 9.1.1.1.5.Port "COM8" does not exist
- 9.1.1.1.6.Serial port initialization error
- 9.1.1.1.7.Port "COM4" does not exist
- 9.1.1.1.8.Cannot connect to Data Recorder
- 9.1.1.1.9.Password error
- 9.1.1.1.10."Print error" prompt box
- 9.2Problems with Translational Rotation Platform
- 9.3Problems with Capsule
- 9.4Problems with Data Recorder
- 9.5Problems with Locator
- 9.1ESNavi Error Messages
- Chapter 10:TECHNICAL SPECIFICATIONS
- Chapter 11:SYSTEM LABELING
NaviCam
®
Stomach Capsule User Manual Page | 105
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
indicate current battery capacity. It is 90% when
③④⑤
LEDs are all lit up, more
than 70% when
④⑤
LEDs are lit up, and 40% when
⑤
LED is lit up. If only
⑤
LED is on/off flashing, the remaining battery capacity is very limited. If the Data
Recorder is powered on and being charged, these three LEDs will be on and off
flashing in a cycle manner. If the white LED charge indicator (marked as in
⑥
in
Figure 6- 3) keeps lighting up, it indicates that it is under charging (it will be lit up
once it is charged regardless whether the Data Recorder is powered on). The white
LED will automatically turn off when fully charged.
Warning
The Data Recorder should be charged only with provided charger.
Charging can only be performed by the physician/operator. Patients shall not
have access to the adaptor.
The USB interface should only be connected to the PC of console.
6.2 Usage of Locator
6.2.1 Operation Panel
Figure 7-1 shows functional keys and indicators of the Locator.
Figure 7-1 Operation Panel