User's Manual
Table Of Contents
- Chapter 1:INTRODUCTION
- Chapter 2:INDICATIONS, CONTRAINDICATIONS AND WARNINGS
- Chapter 3:WIRING AND CONTROL PANEL OF CONSOLE
- Chapter 4:OPERATION INSTRUCTION FOR ESNAVI SOFTWARE
- 4.1Installing Software
- 4.2Uninstalling Software
- 4.3Basic Operation of Software
- Chapter 5:USAGE OF CAPSULE
- Chapter 6:USAGE OF DATA RECORDER
- Chapter 7:INSTALLATION AND TRAINING
- Chapter 8:MAINTENANCE
- Chapter 9:WARNINGS, CAUTIONS AND TROUBLESHOOTING
- 9.1ESNavi Error Messages
- 9.1.1.1.1.Check if camera is connected firmly
- 9.1.1.1.2.Equipment initialization error
- 9.1.1.1.3.Port "COM7" does not exist
- 9.1.1.1.4.Video monitoring connection fails
- 9.1.1.1.5.Port "COM8" does not exist
- 9.1.1.1.6.Serial port initialization error
- 9.1.1.1.7.Port "COM4" does not exist
- 9.1.1.1.8.Cannot connect to Data Recorder
- 9.1.1.1.9.Password error
- 9.1.1.1.10."Print error" prompt box
- 9.2Problems with Translational Rotation Platform
- 9.3Problems with Capsule
- 9.4Problems with Data Recorder
- 9.5Problems with Locator
- 9.1ESNavi Error Messages
- Chapter 10:TECHNICAL SPECIFICATIONS
- Chapter 11:SYSTEM LABELING
NaviCam
®
Stomach Capsule User Manual Page | 28
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
Figure 4- 16 Image Scaling Function
2. If you wish to restore the original size of the displayed image without any zoom in
or out, double click the image display area.
4.3.2.4. Data Recorder Status Indicator
1. Rtdisply view of the two LED status indicators in the lower right corner are as
shown in Figure 4-17. The left indicates the work status (Green), and the right
indicates warning (red).
Figure 4- 17 Status Indicators
2. Meaning of the status indicators:
(1) Green LED flashes and red LED does not light indicate image data from capsule
endoscope has been received.
(2) Green LED does not light and red LED lights indicate no signal from capsule
endoscope has been received.
(3) Green LED does not light and red LED flashes indicate capsule endoscope signal
has been received, but no image data has been received.
4.3.2.5. Search Capsule Endoscope
If the capsule endoscope has been turned on, and capsule current channel is unclear,
then the search capsule endoscope function can be used by:
1. First turn on the capsule endoscope with Locator for more than 3 seconds.