User's Manual

Table Of Contents
NaviCam
®
Stomach Capsule User Manual Page | 21
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
4.3 Basic Operation of Software
Basic flow of the software when performing a procedure:
Step one - enter basic and patient information.
Step two - save the information to the Data Recorder.
Step three - turn on the capsule endoscope.
Step four - observe and adjust capsule endoscope working status by viewing real-time
display monitor.
Step five - examine the patient.
Step six - export image data captured by the capsule endoscope to computer by means
of data export.
Step seven - view and annotate the image data.
Step eight - generate patient report. If there are several cases, the data of the cases can
be managed on the case management interface. For details, refer to software use
flowchart as shown in Figure 4- 9.
Figure 4- 9 Software Use Flowchart
4.3.1 Patient Check-In
1. Open the ESNavi software to access software registration interface as shown in
Figure 4- 10. User feedback Serial No. and input the obtained registration license
No. to register in the software. ESNavi software cannot be run if registration is
not entered successfully.