User's Manual
Table Of Contents
- Chapter 1:INTRODUCTION
- Chapter 2:INDICATIONS, CONTRAINDICATIONS AND WARNINGS
- Chapter 3:WIRING AND CONTROL PANEL OF CONSOLE
- Chapter 4:OPERATION INSTRUCTION FOR ESNAVI SOFTWARE
- 4.1Installing Software
- 4.2Uninstalling Software
- 4.3Basic Operation of Software
- Chapter 5:USAGE OF CAPSULE
- Chapter 6:USAGE OF DATA RECORDER
- Chapter 7:INSTALLATION AND TRAINING
- Chapter 8:MAINTENANCE
- Chapter 9:WARNINGS, CAUTIONS AND TROUBLESHOOTING
- 9.1ESNavi Error Messages
- 9.1.1.1.1.Check if camera is connected firmly
- 9.1.1.1.2.Equipment initialization error
- 9.1.1.1.3.Port "COM7" does not exist
- 9.1.1.1.4.Video monitoring connection fails
- 9.1.1.1.5.Port "COM8" does not exist
- 9.1.1.1.6.Serial port initialization error
- 9.1.1.1.7.Port "COM4" does not exist
- 9.1.1.1.8.Cannot connect to Data Recorder
- 9.1.1.1.9.Password error
- 9.1.1.1.10."Print error" prompt box
- 9.2Problems with Translational Rotation Platform
- 9.3Problems with Capsule
- 9.4Problems with Data Recorder
- 9.5Problems with Locator
- 9.1ESNavi Error Messages
- Chapter 10:TECHNICAL SPECIFICATIONS
- Chapter 11:SYSTEM LABELING
NaviCam
®
Stomach Capsule User Manual Page | 112
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
Improper replacement of the fuse can induce harm to both the personnel and the
device.
Please follow the fuse replacing instruction to safely replace the fuse.
8.2 Inspection of Magnet strength
The service life of the Controller’s magnetic ball is 6 years.
Check that the maximum magnetic field value at the lowest 30 mm. of the magnetic
head is 189.8 55.0 mT. If it does not meet this value, the magnet of the Controller
may need to be replaced or serviced.
8.3 System Cleaning
Your facility should provide a bed cover that meet medical standards. To avoid cross
contamination between patients, please replace with a new cover for each patient.
Once a month wipe the examination bed surface with cleaning cloth lightly dipped in
70% isopropyl alcohol and let it dry for 15 minutes.
Once a month, in power off condition, wipe the instrument surface with cleaning cloth
lightly dipped in 70% isopropyl alcohol and let it dry for 15 minutes. If cleaning of
internal parts of the equipment is required, only qualified personnel may clean the
inside panels. ALWAYS make sure it is powered off. Outer casing may be cleaned by
air blower to remove dust.
Preventive inspection and maintenance of the examination bed surface, frame and
electric equipment should be conducted at least once yearly by a professional
technician. Check for loose fasteners or other factors that may cause injury to patients
or operators.
8.4 Screw Lubrication
Lubricate the ball screws every three months, then start up the equipment and operate
the equipment for 2-3 trips in XYZ directions.
Note
Reinstall the dust cover after lubrication, close the equipment back cover
and lock it.