User's Manual
Table Of Contents
- Chapter 1:INTRODUCTION
- Chapter 2:INDICATIONS, CONTRAINDICATIONS AND WARNINGS
- Chapter 3:WIRING AND CONTROL PANEL OF CONSOLE
- Chapter 4:OPERATION INSTRUCTION FOR ESNAVI SOFTWARE
- 4.1Installing Software
- 4.2Uninstalling Software
- 4.3Basic Operation of Software
- Chapter 5:USAGE OF CAPSULE
- Chapter 6:USAGE OF DATA RECORDER
- Chapter 7:INSTALLATION AND TRAINING
- Chapter 8:MAINTENANCE
- Chapter 9:WARNINGS, CAUTIONS AND TROUBLESHOOTING
- 9.1ESNavi Error Messages
- 9.1.1.1.1.Check if camera is connected firmly
- 9.1.1.1.2.Equipment initialization error
- 9.1.1.1.3.Port "COM7" does not exist
- 9.1.1.1.4.Video monitoring connection fails
- 9.1.1.1.5.Port "COM8" does not exist
- 9.1.1.1.6.Serial port initialization error
- 9.1.1.1.7.Port "COM4" does not exist
- 9.1.1.1.8.Cannot connect to Data Recorder
- 9.1.1.1.9.Password error
- 9.1.1.1.10."Print error" prompt box
- 9.2Problems with Translational Rotation Platform
- 9.3Problems with Capsule
- 9.4Problems with Data Recorder
- 9.5Problems with Locator
- 9.1ESNavi Error Messages
- Chapter 10:TECHNICAL SPECIFICATIONS
- Chapter 11:SYSTEM LABELING
NaviCam
®
Stomach Capsule User Manual Page | 111
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
CHAPTER 8: MAINTENANCE
Maintenance of the equipment should be completed by qualified personnel at least
once a year.AnX Robotica Corp. will make available on request circuit diagrams,
component part lists, descriptions, calibration instructions, or other information that
will assist service personnel to repair the device. Do not attempt to make any
modifications to the device. Any modification during the actual service life require
re-evaluation of safety and EMC.
The following routine inspection and cleaning is required:
Power Inspection
Regularly check power supply of the System for damage, cracking or wear. Do not
use the equipment when power voltage is out of specified range for the equipment
120VAC, 60Hz.
8.1 Replacement of Fuse
There are two fuses with fast-acting and high breaking capacity features. Each
specified as 125VAC 6A, and the dimension is Φ5mm in diameter, 20mm in length.
The two fuses are located near the power input terminal of the equipment (see Figure
9-1). Please use a fuse removal tool to remove the fuse caps and replace the related
fuse. AnX Robotica recommends that the fuses by replaced annually.
Figure 9- 1: Schematic Diagram of fuse replacement
Warning
Replacing of the fuse can only be done by qualified personnel.