User's Manual

Table Of Contents
NaviCam
®
Stomach Capsule User Manual Page | 105
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
indicate current battery capacity. It is 90% when
③④⑤
LEDs are all lit up, more
than 70% when
④⑤
LEDs are lit up, and 40% when
LED is lit up. If only
LED is on/off flashing, the remaining battery capacity is very limited. If the Data
Recorder is powered on and being charged, these three LEDs will be on and off
flashing in a cycle manner. If the white LED charge indicator (marked as in
in
Figure 6- 3) keeps lighting up, it indicates that it is under charging (it will be lit up
once it is charged regardless whether the Data Recorder is powered on). The white
LED will automatically turn off when fully charged.
Warning
The Data Recorder should be charged only with provided charger.
Charging can only be performed by the physician/operator. Patients shall not
have access to the adaptor.
The USB interface should only be connected to the PC of console.
6.2 Usage of Locator
6.2.1 Operation Panel
Figure 7-1 shows functional keys and indicators of the Locator.
Figure 7-1 Operation Panel