User's Manual
Table Of Contents
- Chapter 1:INTRODUCTION
- Chapter 2:INDICATIONS, CONTRAINDICATIONS AND WARNINGS
- Chapter 3:WIRING AND CONTROL PANEL OF CONSOLE
- Chapter 4:OPERATION INSTRUCTION FOR ESNAVI SOFTWARE
- 4.1Installing Software
- 4.2Uninstalling Software
- 4.3Basic Operation of Software
- Chapter 5:USAGE OF CAPSULE
- Chapter 6:USAGE OF DATA RECORDER
- Chapter 7:INSTALLATION AND TRAINING
- Chapter 8:MAINTENANCE
- Chapter 9:WARNINGS, CAUTIONS AND TROUBLESHOOTING
- 9.1ESNavi Error Messages
- 9.1.1.1.1.Check if camera is connected firmly
- 9.1.1.1.2.Equipment initialization error
- 9.1.1.1.3.Port "COM7" does not exist
- 9.1.1.1.4.Video monitoring connection fails
- 9.1.1.1.5.Port "COM8" does not exist
- 9.1.1.1.6.Serial port initialization error
- 9.1.1.1.7.Port "COM4" does not exist
- 9.1.1.1.8.Cannot connect to Data Recorder
- 9.1.1.1.9.Password error
- 9.1.1.1.10."Print error" prompt box
- 9.2Problems with Translational Rotation Platform
- 9.3Problems with Capsule
- 9.4Problems with Data Recorder
- 9.5Problems with Locator
- 9.1ESNavi Error Messages
- Chapter 10:TECHNICAL SPECIFICATIONS
- Chapter 11:SYSTEM LABELING
NaviCam
®
Stomach Capsule User Manual Page | 103
AnX Robotica Corp. UM-NCES-ST-01 Rev 1
CHAPTER 6: USAGE OF DATA RECORDER
6.1 AKR-1 Data Recorder
6.1.1 Installation and Usage
The Data Recorder is equipped with 14 sensor arrays, as presented in Figure 6- 1.
Figure 6- 1 Data Recorder (AKR-1)
When in use, the Data Recorder and sensor arrays are embedded into the
corresponding examination cloth as shown in Figure 6- 2. As tested, the working
temperature of the data recorder can reach 42℃ tested under 40℃ ambient
temperature. Since the data recorder does not contact patient skin directly, the
temperature will not cause burn or other hazard to patients and operators, and the
contact duration can be longer than 10 minutes.
Figure 6- 2 Placement of Sensors