User Manual
11
7. Software Control Panel
1. Device S/N
Shows which device is controlled. If you have multiple devices connected to your PC/ MAC
computer (or different computers wired in the same Local Area Network) you can select
which one to control.
2. Calibration Mode
Enters the calibration mode.
Ensure atomic signal is fed to the Atomic Input, the Atomic Clock is warmed up and the
LiveClock Atomic Clock Indicator is lit.
1. Click on calibration and the device will enter calibration mode
2. The indicator below will show a number which will stabilize with time. Wait for the
number to become stable. This number represents the error of the LiveClock clock related
to the Atomic Clock. To compensate for this error press the “Calibrate” button.
1
5
9
6
7
8
2 3 4