Presets Saving
Enables you to save the current settings on a preset.
Factory Reset
Resets the device to the factory settings, note that all settings and saved presets will be lost.
10M Calibration
Enters calibration mode which will enable you to calibrate the Goliath to a 10M.
Ensure atomic signal is fed to the Atomic Input, the Atomic Clock is warmed up and the Atomic
Indicator on the touchscreen is lit.
1. Select 10M calibration from the menu and the device will enter calibration mode
2. The indicator below will show a temperature number which will stabilize with time. When
the number is stable and the 10M is connected LOCKED will appear on the screen. Wait
until the number of “clock errors” are displayed. Click Done when the error number gets
stabilized to nish the calibration.
3. Repeat the procedure as many times as necessary so that the number becomes as
close as possible to Zero and relatively stable (number may have small deviations in a
few seconds).
Device Info
Displays the serial number, hardware revision and rmware version of the device.
Touch Screen Calibration
Will enter the touch screen calibration mode. Follow the instructions on the screen to re-calibrate
the screen.
Lock
Indicates whether the Goliath is locked to an incoming signal or not.
Power
Powers off the device, if the Goliath is in standby mode, a power button like this will appear:
Tap it to switch the device on.