User guide

High Accuracy Power Meter (Option 19) Error Messages
The following error messages could be displayed on the Site Master when using the High
Accuracy Power Meter mode with Power Sensor PSN50. If any error condition persists,
contact your local Anritsu Service Center.
General Error Messages
A listing of General Error Messages is provided below.
2-22
Chapter 2 Functions and Operations
Error Message Description
Warning! Power
Supply Error
Verify that the supply is connected properly.
Warning! RF Power
Level is too high
The specified upper measurement range is +20 dBm. Do not exceed
the specification.
Warning! Sensor
not zeroed properly.
Zero sensor again
The sensor must be zeroed with nothing connected to it.
Warning! Specified
temperature range
(0 to 50°C) exceeded
The PSN50 sensor temperature range is 0 to 50°C.
Warning! Tempera-
ture has changed.
Zero sensor again
The temperature changed more than the allowable limit after the sensor
was zeroed.
Table 2-5. Option 19 Error Messages
Error Message
Description
CAL
Incomplete
A complete open, short, and load calibration must be performed before
calibration can be turned on.
Dist Requires
F1<F2
Valid distance to fault plots require a non-zero frequency span.
Invalid Sweep Data The sweep data is invalid.
Use Options menu
to select a printer
Attempting to print a display with no printer selected. Select a printer,
then retry.
Flexcal is not suit-
able for this case,
Change To OSL
CAL mode
(SYS-Options)
The cable is too long or too reflective for FlexCal to work. OSL cal
must be used instead.
Cannot zero
input signal too high
Attempting to perform a Power Meter zero adjust function with an in-
put of greater than –20 dBm.
Table 2-6. General Error Messages