User guide
Table Of Contents
- Title Page
- Table of Contents
- Chapter 1 General Information
- Chapter 2 Functions and Operations
- Chapter 3 Getting Started
- Chapter 4 Cable and Antenna Analyzer Measurements
- Chapter 5 Power Monitor Mode
- Chapter 6 High Accuracy Power Meter
- Chapter 7 Handheld Software Tools
- Chapter 8 Master Software Tools
- Appendix A Windowing
- Index

Performance Specifications
Performance specifications are provided in Table 1-1. All specifications apply when cali
-
brated at ambient temperature after a five minute warm up. Typical values are given for ref-
erence, and are not guaranteed.
1-7
Chapter 1 General Information
Cable and Antenna Analyzer
Frequency Range: 25 MHz to 4000 MHz
Frequency Accuracy: £±50 ppm @ +25°C
Frequency Resolution: 1 kHz (CW On)
100 kHz (CW Off)
Output Power: 0 dBm (typical)
Immunity to Interfering Signals: on-channel +17 dBm
on-frequency –5 dBm
Measurement speed: £ 2.5 msec / data point (CW ON)
Number of data points: 130 or 259 or 517
Return Loss:
Range: 0.00 to 60.00 dB
Resolution: 0.01 dB
VSWR:
Range: 1.00 to 65.00
Resolution: 0.01
Cable Loss:
Range: 0.00 to 30.00 dB
Resolution: 0.01 dB
Measurement Accuracy: > 42 dB corrected directivity after calibration
Distance-To-Fault
Vertical Range:
Return Loss: 0.00 to 60.00 dB
VSWR: 1.00 to 65.00
Horizontal Range: 0 to (# of data pts –1) x Resolution to a maximum of
1497m (4911 ft)
# of data pts = 130 or 259 or 517
Horizontal Resolution (rectangular windowing):
Resolution (meters) = (1.5 x 10
8
) x (Vp)/DF
Where Vp is the relative propagation velocity of the cable and DF is the stop frequency minus the
start frequency (in Hz).
Table 1-1. Performance Specifications (1 of 3)