Specifications

15. Select:
MARKER READOUT
16. Use the soft key to turn on Marker 1. Then use the numeric data en
-
try key to change POINT to 25.
17. Press the Marker key again.
18. Select the SCREEN DISPLAY soft key to ON.
19. Zero and calibrate the power sensor.
20. Set the power sensor CAL FACTOR for the frequency being mea
-
sured.
21. Connect the power sensor to the BNC male end of the adapter con
-
nected to the 43KC-10 fixed attenuator (Figure 3-5).
22. Press the Power key and select SOURCE 1 SETUP. Adjust SOURCE
1 POWER so that the power meter reads -15 dBm ±0.1 dB.
23. Disconnect the power sensor from the BNC male end of the adapter
that is connected to the 43KC-10 fixed attenuator (Figure 3-6).
RECEIVER MAGNITUDE DISPLAY LINEARITY TEST (MS462XC) PERFORMANCE
TESTS
3-20 MS462XX MM
DataEntry
Power
Keyboard
Port3
Channels
Enhancement
Stimulus
System
Port1
Port2
ProbePower
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
m
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch1
Ch2
Ch4
Ch3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save/
Recall
0
7
4
Display
Clr
Print
MS462XC Front Panel
Anritsu Power Meter
Power Sensor
Coaxial Cable
SMA (f) to BNC (m)
Adapter
Port 1
*Connect dashed lines as directed by procedure
*
43KC-10
Attenuator
Cal / Zero
ML 2408A
Power Meter
ON / OFF
Sensor
Channel
Trigger
System
CALIBRATOR
0
1
4
7
2
5
8
+/-
3
6
9
CLR
A
B
Figure 3-5. Test Equipment Setup for Receiver Magnitude Display Linearity Test Calibration Setup 1