Specifications

8. Select the NEXT FREQUENCY soft key, then enter the following fre
-
quencies:
q
MS4622C—1 GHz, 2 GHz, and 3 GHz
q
MS4623C—1 GHz, 2 GHz, 4 GHz, and 6 GHz
q
MS4624C—1 GHz, 2 GHz, 4 GHz, 6 GHz and 9 GHz
9. Select:
RETURN
RETURN to exit this setup menu.
10. Select the C.W. MODE soft key to turn C.W. mode ON.
11. Change the C.W. frequency to 50 MHz.
12. Press the Config key and select the DATA POINTS soft key. Change
POINTS DRAWN IN C.W. to 51 POINT(S).
13. Press the Meas key and select USER DEFINED.
14. Select:
CHANGE RATIO
b1 (Ta)
1 (UNITY)
15. Press the S11/USER 1 soft key to select USER 1.
16. Press the Marker key.
17. Select:
MARKER READOUT
18. Select the marker soft key to turn on Marker 1, then use the nu
-
meric data entry key to change POINT to 25.
19. Press the Marker key again.
20. Select the SCREEN DISPLAY soft key to ON.
21. Zero and calibration the power sensor.
22. Set the power sensor CAL FACTOR for the frequency being mea
-
sured; that is, 50 MHz, 1 GHz etc.
23. Connect the power sensor to the unterminated end of the 43KC-10
fixed attenuator with cable linked to Port 1.
24. Press the Power key and select SOURCE 1 SETUP. Adjust SOURCE
1 POWER so that the power meter reads –10 dBm ±0.01 dB.
PERFORMANCE TESTS COMPRESSION LEVEL TEST (MS462XC)
MS462XX MM 3-17
NOTE
Select the INSERT
NEXT FREQUENCY
soft key to store the
entered frequency
value to memory.