Specifications

3-7 COMPRESSION
LEVEL TEST
(MS462XC)
This test verifies the compression magnitude levels of the MS462XC Vec
-
tor Network Measurement System receiver.
Test Setup: Set up the equipment as shown below and allow the instruments to
warm-up for at least one hour.
Procedure:
1. Press the Default key, then the 0 key to reset the instrument.
2. Press the Ch 3 key, then the Display key.
3. Select:
DISPLAY MODE
SINGLE CHANNEL
RETURN
GRAPH TYPE
LOG MAGNITUDE
4. Press the Freq key.
5. Select:
DISCRETE FILL
INDIVIDUAL FREQ INSERT
6. Select AUTO INCR to turn this function off.
7. Select the NEXT FREQUENCY soft key. Enter 50 MHz and select the
INSERT NEXT FREQUENCY soft key.
COMPRESSION LEVEL TEST (MS462XC) PERFORMANCE TESTS
3-16 MS462XX MM
DataEntry
Power
Keyboard
Port 3
Channels
Enhancement
Stimulus
System
Port 1
Port 2
Probe Power
1
8
5
2
9
6
3
-
.
G/ns/m
M/ s/cm
m
k/ms/mm
X1
Enter
Local
Appl
Meas
Marker
Hold
Ch 1
Ch 2
Ch 4
Ch 3
Freq
Sweep
Power
Config
Seq
Cal
Avg
Utility
Hard
Copy
Default
Save/
Recall
0
7
4
Display
Clr
Print
Noise
In
Noise
+28V
NOOPERATOR SERVICE-
ABLEPARTS INSIDE.
REFERSERVICING TO
QUALIFIEDPERSONNEL.
WARNING
CAUTION
FORCONTINUED FIRE
PROTECTIONREPLACE
ONLYWITH SPECIFIED
TYPEAND RATED FUSE.
IEEE488.2
GPIB
Dedicated
GPIB
SerialPort
Ethernet
VGA
SCSI-2
Ext
Analog
Output
Ext
Trigger
LPTLine Printer
ExtI/O
ExtAnalog
In
Ext
Source
Ext10 MHz
In
47-440Hz
85-264VAC
REPLACEFUSE ONLY WITH
SAMETYPE AND RATING
CAUTION
DONOT OPERATE
WITHPOWER CORD
UNGROUNDED
N274
A1
B1
A2 B2
MS46XC Rear Panel
MS462XC Front Panel
Anritsu Power Meter
Power Sensor
Coaxial Cable
Port 1
a1
a2 b1
b2
*Connect dashed lines as directed by procedure
43KC-10
Attenuator
*
Cal / Zero
ML2408A
Power Meter
ON/OFF
Sensor
Channel
Trigger
System
CALIBRATOR
0
1
4
7
2
5
8
+/-
3
6
9
CLR
A
B
Figure 3-4. Test Equipment Setup for Compression Test