Specifications

8. Press the BEGIN CAL soft key to start calibration.
9. After the Flat Test Port Power Cal is completed, disconnect the
power sensor from the unterminated end of the 43KC-10 attenuator
and connect the attenuator with the coaxial cable to b1 input on the
rear panel of MS462XC.
10. Press the Avg key and select:
SELECT I.F. BANDWIDTH
I.F.BW10Hz
11. Press the Meas key and select:
USER DEFINED
CHANGE RATIO
b1 (Ta)
1 (UNITY)
12. Press the S21/USER 1 soft key to select USER 1.
13. Press the Display key and select the TRACE MEMORY soft key.
14. Allow two complete sweeps to occur, the select:
STORE DATA TO MEMORY
DATA (/) MEMORY
RETURN
SCALE
15. Disconnect the 43KC-10 attenuator with the cable from the receiver
input. Then connect a termination to the receiver input.
16. Change the REFERENCE VALUE to - 97 dB.
17. Allow one full sweep to occur, then press the Hold key.
18. Verify that the trace is below the REFERENCE VALUE.
19. Press the Meas key and select:
USER DEFINED
CHANGE RATIO
b2 (Tb)
1 (UNITY)
20. Remove the 43K-10 Attenuator with the coaxial cable from b1 input
and connect to b2 input.
21. Repeat Steps 13 to 18.
PERFORMANCE TESTS SYSTEM DYNAMIC RANGE TEST (MS462XC)
MS462XX MM 3-15