Specifications

20. Press the Display key and select:
GRAPH TYPE
LOG MAGNITUDE
RETURN
SCALE
RESOLUTION
21. Enter 10 and press the X1 key.
22. Select REFERENCE VALUE and enter -70, then press the X1 key.
Reference line 7 and select RETURN.
23. Select:
DISPLAY MODE
SINGLE CHANNEL
24. Press the Meas key and select:
S21, TRANS b2/a1
25. Press the Marker key and select:
DISPLAY MARKERS ON
READOUT MARKERS
26. Set the markers as follows:
MARKER 1: 10MHz
MARKER 2: 400 GHz
MARKER 3: 3 GHz
MARKER 4: 6 GHz
MARKER 5: 9 GHz
MARKER 6: Any frequency
27. Allow the instrument to complete a full sweep. Using the knob to
move Marker 6, view the Marker 6 readout and find the highest
point between subsequent markers. Ensure that the values of the
highest points are below the values shown in Table 3-2.
28. If the instrument has two ports, the Dynamic Range test is finished.
29. If the instrument has three or four ports, calibrate for S13 as fol
-
lows:
30. Press the Cal key, then select:
MANUAL CAL
3 PORT
NEXT CAL STEP
SYSTEM DYNAMIC RANGE TEST (MS462XA/B/D) PERFORMANCE TESTS
3-12 MS462XX MM