Specifications

9. Ensure the correct test port connector types are selected. If not, se
-
lect the port connection displayed and choose the correct type from
the list.
10. Select START CAL and continue to use the default settings for the
calibration.
11. Before measuring the isolation devices (50 W terminations), press
the Avg key.
12. Select DATA AVERAGE and enter 10 (followed by the X1 key).
13. Select:
I.F. BANDWIDTH
I.F.BW10Hz
14. Install the isolation devices (50 W terminations) from the Cal Kit to
Port 1 and Port 2.
15. Select the MEASURE BOTH PORTS soft key.
16. Continue the calibration by following the on-screen directions and
by selecting the following soft keys until the calibration is complete:
NEXT CAL STEP
MEASURE BOTH PORTS or
MEASURE DEVICE(S)
17. The calibration is complete when the following message is displayed
(press the Enter key):
PRESS <ENTER> TO PROCEED
18. Install the 50 ohm terminations from the Cal Kit to Port 1 and
Port 2.
19. Press the Avg key and set the number of averages to 10 and the
I.F. BW to 10 Hz by selecting:
AVERAGE ON
DATA AVERAGE
10
X1
SELECT I.F. BANDWIDTH
I.F.BW10Hz
PERFORMANCE TESTS SYSTEM DYNAMIC RANGE TEST (MS462XA/B/D)
MS462XX MM 3-11