Specifications

6. Change the user defined ratio by selecting:
CHANGE RATIO
MORE
b2 (Tb) as the numerator
1 (UNITY) as the denominator
7. Ensure thta the minimum level of b2/1 is above the level shown in
Table 3-1.
NON-RATIO PARAMETER TEST (MS462XA/C/D) PERFORMANCE TESTS
3-8 MS462XX MM
Model
Beginning
S-Parameter
Device to
Install
User Defined
Ratio
Minimum
Level
MS4622A
S21
Thru from
Port 1 to
Port 2
a1/1 -6 dB
S21 b2/1 -12 dB
MS4623A S21 a1/1 -8 dB
S21 b2/1 -13 dB
MS4624A
S21 a1/1 -15 dB
S21 b2/1 -15 dB
MS4622B
S21
Thru from
Port 1 to
Port 2
a1/1 -7 dB
S21 b2/1 -18 dB
S12 a2/1 -7 dB
S12 b1/1 -18 dB
S33
Short on
Port 3
a3/1 -7 dB
S33 b3/1 -18 dB
MS4623B
S21
Thru from
Port 1 to
Port 2
a1/1 -8 dB
S21 b2/1 -15 dB
S12 a2/1 -8 dB
S12 b1/1 -10 dB
S33
Short on
Port 3
a3/1 -8 dB
S33 b3/1 -11 dB
MS4624B
S21
Thru from
Port 1 to
Port 2
a1/1 -15 dB
S21 b2/1 -20 dB
S12 a2/1 -15 dB
S12 b1/1 -15 dB
S33
Short on
Port 3
a3/1 -15 dB
S33 b3/1 -15 dB
MS4624D
S21
Thru from
Port 1 to
Port 2
a1/1 -15 dB
S21 b2/1 -15 dB
S12 a2/1 -15 dB
S12 b1/1 -15 dB
S33
Short on
Port 3
a3/1 -15 dB
S33 b3/1 -20 dB
S44
Short on
Port 4
a4/1 -15 dB
S44 b4/1 -20 dB
Table 3-1. Minimum Amplitude Specifications