7. Verify that the instrument passes this test.
8. If the instrument is a MS462XB/C with Option 3 installed or a
MS462XD, press the SELECT SOURCE soft key to select source 2.
9. Connect the power sensor to Port 3 and select the START VERIFICA
-
TION soft key.
10. Verify that the instrument passes this test.
11. Select the menu soft key as follows to exit the TROUBLESHOOTING
mode:
RETURN
RETURN
FINISHED, RECOVER FROM TROUBLESHOOTING
VERIFYING OUTPUT POWER ACCURACY PERFORMANCE TESTS
3-6 MS462XX MM
NOTE
The worst case test
result will be
displayed on the
screen when the
instrument fails this
test only. The failed
test result is also
recorded in the
Service Log.
CAUTION
The TROUBLESHOOTING function must be properly exited to restore
normal system operation.