Specifications
Instrument Critical Specifications
Recommended
Manufacturer or Model
Application
Code
Power Meter
and Sensor
Power Meter:
Power Range –70 to
+20 dBm
GPIB controllable
Power Sensor:
Frequency Range
10 MHz to 18 GHz
Power Range –70 to
+20 dBm
Anritsu ML2430A Series
Anritsu MA2472A
O, A, P
Digital
Multimeter
Resolution: 4 ½ digits
DC Accuracy: 0.1 %
AC Accuracy: 0.1 %
Any T
Oscilloscope Bandwidth :
DC to 100 MHz
Sensitivity: 2 mV
Horiz. Sensitivity:
50 ns/division
Tektronix Inc., Model 2445 T
Power Supply Voltage: +12V
Current: 100 mA min.
Any O
Power
Amplifier
Gain: 10 dB min.
Noise Figure: 5.5 dB
typical
Mini-Circuit
ZJL-4G (4 GHz) APN:60-242
ZJL-6G (6 GHz) APN:60-234
O
Noise Source 15 dB ENR Anritsu NC346B O
Step
Attenuator
With calibration data at
50 MHz
Anritsu MN510C P
Attenuator 10 dB Anritsu 43KC-10 P
Application Codes:
A = Adjustment / Internal Hardware Calibration
O = Operational Testing
P = Performance Verification
T = Troubleshooting
GENERAL INFORMATION RECOMMENDED TEST EQUIPMENT
MS462XX MM 1-9