Specifications

8. If the instrument is fitted with Option 4X, press the Appl key and se
-
lect:
CHANGE APPLICATION SETUP
MEASUREMENT TYPE TRANSMISSION AND REFLECTION
NOISE FIGURE / NOISE FIGURE SETUP
ENR TABLE OPERATIONS
SAVE ENR TABLE
TO FLOPPY DISK
9. Save the following three files to floppy disk:
q
Vendor ENR Table
q
Internal ENR Extension Table
q
External ENR Extension Table
NOTES
Some of the ENR files may have been stored on the hard
disk; however, ENR files are not normally stored on the hard
disk. The External ENR Extension Table file will only exist if it
has been created by the user. To completely backup all the
files on the hard disk, it is essential that all of the previous
steps be completed.
10. Remove the CPU board from the instrument as described in Sec-
tion 7-4.
11. Remove the drops of RTV compound which hold the lithium battery
in place. Gently slide the battery sideways (gently lift up on the re
-
taining clip only a small since excess lifting will reduce contact pres
-
sure on the new battery).
12. Install a new battery and hold it in place with three drops of RTV
compound, as before.
13. Reassemble the instrument.
14. Turn the instrument ON and press UTILITY MENU and select FOR
-
MAT HARD DRIVE.
15. After formatting the hard drive, reload all files from the floppy disk
back onto the hard drive.
7-16 MS462XX MM
CPU ASSEMBLY REMOVAL AND REPLACEMENT