Specifications
Manual
Calibration
Procedure:
1. On the MS462XX, press the UTILITY front panel key.
2. Select the menu soft keys as follows:
DIAGNOSTICS
HARDWARE CAL (SERVICE USE ONLY)
10 MHz CAL MANUAL
5-6 MS462XX MM
10 MHz CALIBRATION PROCEDURE ADJUSTMENTS
Figure 5-2. Automatic 10 MHz Calibration
Anritsu MF241XB
Frequency Counter
Input 1
Figure 5-3. Manual 10 MHz Calibration Setup
Anritsu MF2412B