Specifications

Automatic
Calibration
Procedure:
1. Connect a GPIB cable (Anritsu 2100-2) between the EIP frequency
counter GPIB connector and the MS462XX dedicated GPIB connec-
tor, as shown in Figure 5-1.
NOTE
If the EIP Frequency Counter does not have the High Stabil-
ity internal 10 MHz Reference Time Base option installed or
the EIP Frequency Counter internal 10 MHz Reference Time
Base has not been calibrated, apply a high stability 10 MHz
reference source to the Frequency Counter instead of using
the internal 10 MHz Reference.
2. On the MS462XX, press the UTILITY front panel key.
3. Select the menu soft keys as follows:
DIAGNOSTICS
HARDWARE CAL (SERVICE USE ONLY)
10 MHz CAL AUTOMATIC
4. Connect a coaxial RF cable between the Band 3 Input connector of
the frequency counter and the Port 1 connector of the MS462XX.
5. Press the START CAL soft key to begin the calibration. Refer to Fig
-
ure 5-2 on the following page.
ADJUSTMENTS 10 MHz CALIBRATION PROCEDURE
MS462XX MM 5-5
Dedicated
GPIB
GPIB
Figure 5-1. Automatic 10 MHz Calibration Setup